4-13
TotalAlert Innity™ HTM Medical Gas Notication System
4107 9016 60.01
Figure 42: High Alarm Settings Tab
Figure 43: Zero Oset Tab
Active Alarm Instruction:
To set facility instructions for active alarms, touch
either the GENERAL MESSAGE or OPTION MSG tab.
The alarm defaults to GENERAL MESSAGE. See Sec-
tion 4.6 for how to set up custom instructions.
G = High Alarm Settings Tab
You can remove the high alarm set point or customize
the high alarm set point.
To remove the high alarm set point:
1. Touch the NONE tab.
2. Touch the check on the CONFIRM SETTINGS
CHANGE screen.
To customize the high alarm set point:
1. Touch the ADJUST tab.
2. Enter the new set point.
3. Touch the Save button.
H = Zero Oset Tab
In some cases, facility dial gauges may show slightly
dierent readings than the TotalAlert Innity™ alarm
(e.g., 3.2 BAR versus 3.0 BAR). To match the
TotalAlert Innity™ alarm exactly to a facility dial gauge:
1. Select the oset amount.
2. Touch the Save button to close the screen.
Conguring Additional Gas Badges
Congure additional gas badges by touching the
number on the main conguration screen that
corresponds with the gas badge and repeating the
steps listed above.