Balance System SD (with v4.X software) 57
Figure 7.12. The Motor Control Testing screen.
Performing a Motor Control Test
1. Position the support handles and the display height and tilt for patient comfort.
2. At the Main Menu, touch < Testing>.
3. Touch <Motor Control>. The Patient Setup Information screen is displayed. If this is a new
patient and the training session is to be saved after its completion, the patient’s name and
height must be entered. If the training session does not need to be saved, designate a
height value, which is correlated to the software’s scoring algorithm based on the calculated
Center of Gravity.
4. Touch <Test Options> to advance to the Motor Control Testing Options screen if desired.
• Here the test trial time can be set, the sensory conditions to be tested can be selected,
the number of trials can be entered, and the rest countdown time (i.e., the time between
trials) can be entered.
• To set the Number of Trials, Platform Stability Setting, or Rest Countdown, touch the
appropriate key and enter the setting from the keypad displayed.
• To set the Motor Control Hold Time, use the <"> or <#> keys to select a time, ranging
from .25 to 5 full seconds.
5. Touch <OK> to confirm the selections and return to the Patient Setup screen.