Balance System SD (with v4.X software) 61
Figure 7.15. The Bilateral Comparison Testing screens--one for each leg.
NOTE: To stop a test in progress at any time and return the platform to a locked
position, touch <Stop>.
11. When the first trial is finished, the screen will display Trial 1 Complete,and a ten-second
rest countdown will begin for the second trial. (ten seconds is the default time; this can be
adjusted in the Test Options.) Touch <Collect Data> to begin the second test trial and
continue in the same manner to complete subsequent trials.
12. After completing all trials, a Test Completemessage is displayed. Touch <Results> to
advance to the Bilateral Comparison Test Results screen.
13. At the Bilateral Comparison Test Results screen, touch <Print> to automatically generate a
printed report if desired.
14. To save the test data from the Results screen, touch <Save Results> and touch <OK> in
response to the Save Results for later reporting or export? prompt. The system will display
Save Results Completedafter the results are saved. If a patient identifier was not entered
at the start of the session (i.e., a name or an ID number), it will need to be entered now in
order to save the results of the session. If the test result is not saved and <Back> or