S10/S12/S10A/S12A Patient Monitor User’s Manual
10-6
are as follows:
1. Select the SpO2 parameter area or waveform area to enter the【SpO2】menu.
2. Select【SpO2
Setup】submenu.
3. Switch off【Pitch Tone】.
10.5.6. Setting PI Display
You can switch on or off PI display by following these steps:
1. Select the SpO2 parameter area or waveform area to enter the【SpO2】menu.
2. Select【SpO2Setup】submenu.
3. Set【Display PI】as to【On】or【Off】.
10.6. Setting PR
10.6.1. Setting PR Alarm
You can set PR alarm by following these steps:
1. Select the SpO2 parameter area or waveform area to enter the【SpO2】menu.
2. Select【PR Alarm】submenu.
3. Set alarms as needed.
10.6.2. Setting QRS Volume
If the alarm source is set to PR, the QRS tone is derived from PR measurements. To
set the QRS volume, follow this procedure:
1. Select the SpO2 parameter area or waveform area to enter the【SpO2】menu.
2. Select【PRSetup】submenu.
3. Set【QRSVolume】to the appropriate value.
If the SpO2 value is effective, the monitor also adjusts the QRS tone (Pitch tone)
according to the SpO2 value. For information, see 10.5.5Setting Pitch Tone.
10.6.3. Setting PR Source
Current pulse source is displayed in the PR parameter area. The PR from current pulse
source has the following characteristics:
PR is monitored as system pulse and generates alarms when you select PR as