S10/S12/S10A/S12A Patient Monitor User’s Manual
12-10
12.9. Assisting Venous Puncture
You can use the NIBP cuff to cause sub-diastolic pressure to block the venous blood
vessel and therefore help venous puncture. To assist venous puncture, follow this
procedure:
1. Select the NIBP parameter area.
2. SelectSetupsubmenu
3. SetAssisted Venipuncture Pressure.
4. SelectAssisted Venipunctureat the bottom of the menu.
5. Puncture vein and draw blood sample.
6. SelectNIBP Start/Stopquick key orAssisted Venipuncturebutton to
manually deflate the cuff. When performing a venipuncture, observe the inflation pressure
and the remaining time of the venipuncture in the NIBP parameter area.
12.10. NIBP Maintenance
12.10.1. NIBP Leakage Test
The NIBP leakage test checks the integrity of the system and of the valve. The NIBP
leakage test should be performed once every two years or when you doubt the NIBP
measurements. The NIBP leakage test should be performed by qualified service personnel
only.
12.10.2. NIBP Calibration
The NIBP accuracy test should be performed once every two years or when you
doubt the NIBP measurements. The NIBP accuracy test should be performed by qualified
service personnel only.
12.11. NIBP Troubleshooting
For more information, see D Alarm Information.