S10/S12/S10A/S12A Patient Monitor User’s Manual
47
Item Default Setting
Alarm print OFF
Measurement mode Single
TI source Manual
Injection volume 10cc
Catheter coefficient 0.564
Temperature Unit ºC
C.8 CO2
Item Default Setting
Et CO2 Alarm switch ON
High limit Adult/Pediatric: 50 mmHg
Neonatal: 45 mmHg
Low limit Adult/Pediatric: 25mmHg
Neonatal: 30mmHg
Alarm
priority
MED
Alarm print OFF
FiCO2 Alarm switch ON
High limit 4 mmHg
Alarm
priority
MED
Alarm print OFF
Apnea delay 20s
Waveform Speed 6.25 mm/s
Scale 50 mmHg
Waveform type Draw
Operating Mode Measurement mode
Unit mmHg
Gas temperature 35 ºC

Table of Contents