S10/S12/S10A/S12A Patient Monitor User’s Manual
16-6
16.6. Setting DM
16.6.1. Setting main unit
You can set the display of the DM main unit by the following steps as below:
1. Select the DM parameter area or waveform area to enter DM menu.
2. Set Unitof DM. The selected unit is displayed in the DM parameter area in
the form of a main unit.
16.6.2. Setting unit conversion parameters
In order to ensure the accuracy of the flow rate, you need to set the conversion
parameter between the number of drops and mL. The steps are as below:
1. Select the DM parameter area or waveform area to enter DM menu.
2. Set Drop Per Milliliter of DM. The default value is 20.
16.7. DM examination
The DM module has already been calibrated before leaving factory. Generally, the
user can directly measure it. Please measure and calibrate the DM module when the
following two situations occur:
——After the DM module is used for half a year to one year;
——Clinicians doubt the accuracy of readings;
DM calibration must be completed by maintenance personnel.