S10/S12/S10A/S12A Patient Monitor User’s Manual
12-6
50% to 80% of the limb.
c) Apply the cuff to the patient’s upper arm or leg and make sure the
marking on the cuff matches the artery location. The cuff should fit
snugly, but with enough room for two fingers to be placed between the
cuff and the patient’s arm (on adults), and loosely on neonates with little
or no air present within the cuff. Otherwise it may cause discoloration
and ischemia of the extremities. Make sure that the cuff index line falls
within the range markings on the cuff.
d) Middle of the cuff should be at the level of the right atrium of the heart.
4. Connect the cuff to the air tubing. Avoid compression or restriction of
pressure tubes. Air must pass unrestricted through the tubing.
CAUTION:
A wrong cuff size and a folded or twisted bladder can cause inaccurate
measurements.
Do not touch or apply external pressure against the cuff and air tubing during
NIBP measurement. This may cause inaccurate blood pressure values.
Use care when placing the cuff on an extremity used for monitoring other
patient parameters.
12.7. Starting and Stopping NIBP Measurements
Start and stop NIBP measurement by selecting the NIBP quick key keys or from the
NIBP menu:
Task Via quick key Via NIBP menu
Start a manual
measurement
Select【NIBP Start/Stop】quick key
Select【Start】
NIBP Auto
measurement
Select【NIBP Measure】quick key
Select interval time
Select【Setup】
submenuset【Interval
Time】
NIBPSequence
Select【NIBP Measure】quick key
Select【Sequence】