S10/S12/S10A/S12A Patient Monitor User’s Manual
7-15
7.12. Other Bed Alarm
Enter other bed observation interface, and when the monitored bed monitor has an
alarm triggered, the alarm light and alarm sound are prompted according to the highest
level of all alarms of the current monitor and other bed monitor. You can view and
manage other bed
alarm.
The alarm delay time from the device to other bed is2s.
7.12.1. Other Bed Alarm Reset
You can reset other bed alarm on the monitor. The steps to enable it to reset the bed
alarm are as follows:
1. Select Main Menu quick keyfrom System column to select
Maintenanceinput maintenance passwordEnter.
2. SelectAlarmsubmenuRemote Viewsubmenu.
3. OpenReset Remote Bed’s Alarms.
And then Bed View Settings (bed number) window Reset Remote Alarm
button in the Alarm submenu will be activated. Click on Reset Remote Alarm
button,
other bed alarm will be reset.
CAUTION
Only when the “Alarm Reset By Other Bed” function of the remote monitor is
enabled, can you reset other bed alarm on this monitor
7.12.2. Authorizing the Alarm Reset to Other Devices
Alarms on your monitor can be reset by remote devices if you enable this function.
To do so, follow this procedure:
1. Select Main Menu quick keyfrom System column to select
Maintenanceinput maintenance passwordEnter.
2. SelectAlarmsubmenuRemote Viewsubmenu.
3. OpenAlarm Reset by Other Bedswitch.