S10/S12/S10A/S12A Patient Monitor User’s Manual
8
result
S10A/S12A: At least 4800 groups.
Holographic
waveform
72 hours at maximum. The specific storage time depends on the
waveforms stored and the number of stored waveform.
A.7 Wireless network
Conforming
standards
IEEE802.11a/b/g/n
Operating
frequency
2.4GHz ~ 2.495 GHz, 5.15GHz ~ 5.35GHz, 5.47GHz ~
5.725GHz, 5.725 GHz~5.82GHz
Data security
WPA-PSK、WPA2-PSK
Encryption
AES、TKIP
A.8 Measurement Specifications
The product shall meet the following measurement specifications. If there is no
special indication, the definition of the index shall preferentially refer to the special
standard of the parameter.
A.8.1 ECG
A.8.1.1 Standard
Meet standards of IEC 60601-2-27
Meet standards of IEC 60601-2-25
A.8.1.2 Performance indicators
Electrosurgery
protection
Cut mode: 300W
Coagulate mode: 100W
Recovery time: ≤10s
In compliance with the requirements in clause 202.6.2.101 of