S10/S12/S10A/S12A Patient Monitor User’s Manual
15-11
——【O
2
Concentration】: Set the oxygen concentration. Can choose 0% ~ 100%,
the default value is 16%.
——【N
2
O Concentration】: Set the concentration of N
2
O. Can choose 0% ~ 100%,
the default value is 0%.
WARNING:
Please set the CO
2
corrections according to actual situation, otherwise, the
measured value may be inaccurate and away from actual value.
15.6.6. Operating mode
You can select the operating mode of the CO
2
module according to the actual
situation of the module. The operation steps are as follows:
1. Select the CO
2
parameter area or waveform area to enter the【CO
2
】menu.
2. Select【Setup】submenu.
3. Set【Operation Mode】.
Measure: Select measurement mode is required when measuring with the CO2
module.
Sleep:
When not using CO2 for monitoring, it is recommended to set the CO2
module to sleep mode to increase the life of the CO2 module.
15.7. Entering Intubation Status
When performing intubation during general anesthesia, you can enter the intubation
mode in order to reduce unnecessary alarms. To enter the intubation mode, follow this
procedure:
1. Select the CO2 parameter area or waveform area to enter the【CO
2
】menu.
2. Select【Intubation Status】button.
For the details of the intubation mode, see
7.11 Intubation Status.
15.8. Zeroing
Mainstream and Sidestream / Microflow CO
2
Module
While zeroing is recommended the first time a CO
2
sensor is connected to the monitor,