S10/S12/S10A/S12A Patient Monitor User’s Manual
24-3
C.O. test
CO
2
test Leakage test
Performance test
Module Calibration
Nurse call function test
When you suspect that the function is not normal.
Analog output performance test
Defibrillation synchronization test
Electrical Safety Tests
Select test items based on IEC
60601-1.
1. After repairing or replacing the power
module.
2. Or after the monitor falls.
3. At least once every two years or as required.
Other Tests
Power-on test 1. First installation, or after each reinstallation.
2. After each repair or replacement of main
engine components.
Print test 1. During the first installation.
2. After repairing or replacing the printer.
Recorder check After repairing or replacing the recorder.
Battery
check
Functionality test
1. During the first installation.
2. After replacing the battery
Performance test
Every two months or when the running time of
the battery is significantly shortened.
24.3. Disposing of the Monitor
After the equipment reaches its service life, please dispose of the monitor and its
accessories according to local regulations.
WARNING:
For the disposal of parts and accessories, if there is no corresponding
regulation, local regulations on disposal of hospital waste can be followed.