S10/S12/S10A/S12A Patient Monitor User’s Manual
43
Item Default Setting
Neonatal: 70 mmHg
Low limit Adult: 70 mmHg
Pediatric: 50 mmHg
Neonatal: 25 mmHg
Alarm priority MED
Alarm print OFF
NIBP-sdp Alarm switch ON
High limit 60mmHg
Low limit 20mmHg
Alarm priority MED
Alarm print OFF
Initial pressure Adult: 160 mmHg
Pediatric: 130 mmHg
Neonatal: 100 mmHg
Interval manual
Start Mode Clock
NIBP End tone OFF
Auxiliary venipuncture pressure Adult: 80mmHg
Pediatric: 60mmHg
Neonatal: 40mmHg
Unit mmHg
C.6 IBP
Item Default Setting
IBP-S Alarm switch ON
High limit ART/Ao/UAP/BAP/FAP/LV/P1/P2 arterial pressure
Adult: 160 mmHg
Pediatric: 120 mmHg
Neonatal: 90 mmHg
PA/PAWP:

Table of Contents