S10/S12/S10A/S12A Patient Monitor User’s Manual
21-4
21.3. HL7 Settings
The real-time data, waveforms, and alarms of the monitor can be transmitted to
the hospital’s monitoring system through the HL7 protocol. The operating steps are as
follows:
1. Select Main Menu quick keyfrom System column to select
Maintenanceinput maintenance passwordEnter.
2. Select Network Setup submenu HL7 submenu.
3. Select Parameter, Waveformand Alarmsending function as
required.
From Physiological data column to select monitor as Server
or Client. If select the monitor as Client, set the IP and
Port for the server receiving the real-time data and waveform. And
can set Intervalof data.
From Alarm Data column to select monitor as Server or
Client. If select the monitor as Client
, set the IP and Port
for the server receiving the real-time data and waveform.