S10/S12/S10A/S12A Patient Monitor User’s Manual
24-1
Chapter 24 Maintenance
WARNING:
Hospitals or medical institutions that use monitors should establish perfect
maintenance plans, otherwise may cause monitor failure and unpredictable
consequences, and may endanger personal safety.
The safety checks or maintenance involving any disassembly of the equipment
should be performed by professional servicing personnel. Otherwise, undue
equipment failure and possible health hazards could result.
If necessary, please contact the manufacturer for product circuit diagrams,
parts lists, calibration instructions or other equipment maintenance related
information.
If there is a problem with the monitor, please contact the maintenance
personnel or us.
24.1. Inspection
Before use, after continuous use for 6-12 months, maintenance or upgrade, qualified
maintenance personnel should conduct a comprehensive inspection to ensure the normal
operation and work of the monitor.
Items to be inspected shall include:
The environment and power supply meet the requirements.
There is no mechanical damage to the monitor and accessories.
The power cord has no abrasion and good insulation performance.
Use the specified accessories.
The alarm system functions normally.
The recorder works normally and the recording paper meets the specified
requirements.
The performance of the battery.
Various monitoring functions are in good working condition.
Grounding impedance and leakage current meet the requirements.