EasyManua.ls Logo

Biolight Meditech S10 - C.o

Default Icon
307 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
S10/S12/S10A/S12A Patient Monitor User’s Manual
46
Item Default Setting
CVP /ICP/CPP: cmH2O
Sensitivity MED
Waveform Speed 25 mm/s
Scale type Manual
Scale Upper scale ART /Ao/UAP/BAP/FAP/LV/P1/P2 arterial pressure:
160mmHg
CVP /ICP/RAP/LAP/UVP venous pressure : 20mmHg;
PA/PAWP30mmHg
Lower scale 0mmHg
High-precision cursor
switch
OFF
High-precision cursor ART/ Ao/ UAP/BAP/FAP/LV/P1/P2 arterial pressure:
80mmHg
CVP /ICP/RAP/LAP/UVP venous pressure: 10mmHg
PA/PAWP15mmHg
PPV measurement OFF
PPV source Auto
Waveform
overlay
settup
Left scale 0~160mmHg
Right scale P1/P2: 0~160mmHg
CVP/RAP/LAP/ICP/UVP: 0 ~20 mmHg
Waveform
speed
25 mm/s
Gridlines OFF
C.7 C.O.
Item Default Setting
TB Alarm switch ON
High limit 39.0 ºC
Low limit 36.0 ºC
Alarm priority MED

Table of Contents