S10/S12/S10A/S12A Patient Monitor User’s Manual
7-14
7.11. Intubation Status
The monitor provides the intubation status function during RESP and CO2
monitoring. In this state, the physiological alarms related to RESP and CO2 are shielded,
and the alarm off icon is displayed in the parameter area. During the intubation process of
general anesthesia surgery, the intubation status can be selected to shield unnecessary
alarms.
7.11.1. Entering Intubation Status
To enter the intubation status, choose either of the following ways:
SelectIntubation Statusquick key.
From the bottom of the RESP or CO2 menu to select Intubation
Status button.
Select Main Menu quick keyfrom Alarm column to select
Intubation Status.
7.11.2. Setting Intubation Status Time
The default intubation time is 2 minutes. To change the time, following this
procedure:
1. Select Main Menu quick keyfrom System column to select
Maintenanceinput maintenance passwordEnter.
2. SelectAlarmsubmenuOthersubmenu.
3. SetIntubation Duration.
7.11.3. Exiting the Intubation Status
To exit the intubation status, choose either of the following ways:
SelectIntubation Statusquick key.
From the bottom of the RESP or CO2 menu to select Exit
Intubation Status button.
SelectMain Menuquick keyfromAlarmcolumn to selectExit
Intubation Status.