S10/S12/S10A/S12A Patient Monitor User’s Manual
40
C.2 RESP
Item Default Setting
RR Alarm switch ON
High limit Adult / Pediatric: 30
Neonatal: 100
Low limit Adult / Pediatric: 8
Neonatal: 30
Alarm priority MED
Alarm print OFF
Apnea Alarm switch ON
Alarm priority HIGH, unadjustable
Alarm print OFF
Apnea Delay 20 s
RR Source Auto
Resp Lead RA_LL
Gain ×1
Waveform Speed 6.25 mm/s
Auto Threshold Detection ON
Respiratory anti-drift ON
C.3 SpO2
Item Default Setting
SpO2 Alarm switch ON
High limit
Adult / Pediatric:100%
Neonatal:95%
Low limit
90%
Alarm priority MED
Alarm print OFF
Desat Alarm switch ON