S10/S12/S10A/S12A Patient Monitor User’s Manual
7
Isolated voltage 1500 VAC
Signal type N.C., N.O., Pulse Output (optional);
Rise and drop time 1ms
Defibrillator synchronization signal output
Output impedance 50±10%
Maximum delay 25 ms (from R wave crest to pulse raise)
Amplitude High level: 3.5~5V, providing a maximum of 1 mA output
current;
Low level: 0.5V, receiving a maximum of 5 mA input current.
Pulse width 100ms±10%
Rise and drop time
1ms
Alarm output
Indicates the inherent
delay in determining
the alarm status.
1s
Alarm delay time
from the monitor to
remote equipment
The alarm delay time from the monitor to remote equipment is
2s, measured at the monitor signal output connector.
Alarm signal sound
pressure level range
Within a distance of one meter, the peak volume range of the
audible alarm generated by the equipment is 45 dB (A) ~ 85 dB
(A).
A.6 Data Storage
Trend data
Long trend: 1800hminimum resolution is 10 min
Medium trend: 180h, minimum resolution is 1 min
Short trend: 6h, minimum resolution is 5 second.
Parameter alarm
event
At least 3000 parameter alarm events and associated parameter
waveform at the moment.
ARR events
3000 ARR eventsand the parameter waveform related to the
time of event occurrence.
NIBP measurement
S10/S12: At least 2400 groups.