S10/S12/S10A/S12A Patient Monitor User’s Manual
24-2
If any damage or abnormal phenomenon is found, please do not use the monitor and
immediately contact the medical engineer of the hospital or the maintenance personnel of
the company.
24.2. Maintenance Schedule
The following tasks, except visual inspection, startup detection, touch screen
calibration, battery inspection and recorder inspection, can only be completed by
professional maintenance personnel. Please contact the maintenance personnel in time
when the following maintenance is required. Before testing or maintenance, the equipment
must be cleaned and disinfected.
Check / Maintenance item Recommended frequency
Preventative Maintenance Tests
Visual inspection When first installed or reinstalled.
NIBP test Pressure check 1. If you suspects that the measurement is
incorrect.
2. Following any repairs or replacement of
relevant module.
3. At least once a year.
Leakage test
CO
2
test Leakage test
Performance test
Module Calibration
DM check Module Check 1. The DM module used for half a year to one
year.
2. Clinicians doubt the accuracy of readings.
Performance Tests
ECG test Performance test
1. When you suspect that the measured value
is inaccurate.
2. After the relevant modules are repaired or
replaced.
3. At least once every two years. NIBP and
CO2 modules shall be provided at least
once a year.
Module Calibration
RESP Performance test
SpO
2
test
NIBP test Pressure check
Leakage test
TEMP test
IBP test
Performance test
Pressure zero