S10/S12/S10A/S12A Patient Monitor User’s Manual
4
Operating time
In a new and fully charged battery at (25) ambient temperature,
typical configuration (connected to SpO2; not connected to ECG
and TEMP cable, and NIBP works in an automatic measurement
mode with a time interval of 30 min; Screen brightness is the
factory default value):
S10/S10A: 8h
S12/S12A: 8h
Charge time The monitor is charged to 90% for less than 6 hours and 100% for
less than 8 hours when it is turned off.
When the monitor is turned on, it is charged to 90% for less than
11 hours and 100% for less than 12 hours.
Turn off delay
5 min-15minafter the low battery alarm first occurs
A.5 Hardware Specifications
A.5.1 Display
Host display
Type Color TFT LCD
S10/S10A
Sizediagonal
10.4 inch
Resolution 800×600 pixels
S12/S12A
Sizediagonal
12.1 inch
Resolution 800×600 pixels
External display
Type TFT display
A.5.2 Recorder
Type
BTR50S thermal dot array
Paper width
50 mm±1mm
Recording speed
12.5 mm/s, 25 mm/s, 50 mm/s
Recording waveform
Maximum 3 tracks