S10/S12/S10A/S12A Patient Monitor User’s Manual
2
A.2 Environmental Specifications
Operating
temperature
Mainframe
0℃~40
CO2 module (external)
5℃~40
Operating humidity
15%95%non-condensing
Operating
atmospheric
pressure
Mainframe
57kPa107.4kPa
CO2 module (external)
86 kPa106 kPa
Transportation
and storage
temperature
Mainframe
-20℃~+60
CO2 module (external)
-40℃~+70
Transportation
and storage
humidity
Mainframe
10%95%non-condensing
CO2 module (external)
<90%non-condensing
Transportation
and storage
atmospheric
pressure
Mainframe
16.0kPa107.4kPa
CO2 module (external)
86 kPa106 kPa
CAUTION:
The equipment must be used under the specified environmental specifications,
otherwise it will not meet the technical specifications claimed in this manual
and may lead to unexpected consequences such as equipment damage. If the
performance of the equipment changes due to aging or environmental
conditions, please contact the maintenance personnel.
A.3 Physical Specifications
Model Weight
Size (W×H×D)
Remark
S10/S10A <4kg 168mm* 288mm* 236mm
Including screen, stationary
parameter module, a lithium
battery, a recorder, without

Table of Contents