7.12.3. Switching Off the Remote Device Disconnection Alarm
The monitor can provide an alarm if remote devices are disconnected. By default, the
function is enabled. To disable the alarm, follow this procedure:
1. Select Main Menu quick keyfrom System column to select
Maintenanceinput maintenance passwordEnter.
2. SelectAlarmsubmenuRemote Viewsubmenu.
3. Switch offRemote Disconnected Alarm.
7.13. Detecting Alarm
The monitor automatically performs a self-test at startup. Check that the alarm lamp
illuminates, one after the other, in red, yellow, and cyan, and that an alarm tone is heard.
This indicates that the visible and audible alarm indicators functions correctly.
To further test individual measurement alarms, perform measurements on yourself or
using a simulator. Adjust alarm limits and check that appropriate alarm behavior is
observed
7.14. Actions When an Alarm Occurs
When an alarm occurs, please refer to the following steps to take proper actions:
1. Check the patient’s condition.
2. Confirm the alarming parameter or alarm category.
3. Identify the source of the alarm.
4. Take proper action to eliminate the alarm condition.
5. Check if the alarm is eliminated.
For more information, please refer to D Alarm Information.