S10/S12/S10A/S12A Patient Monitor User’s Manual
7-10
After reaching the alarm pause time, the monitor will automatically exit the alarm
pause state. You can also click 【Alarm Pause】quick key to manually cancel the alarm
pause.
7.7.1. Setting Alarm Pause Time
Alarm pause time can set to :【1min】, 【2min】, 【3min】and【Permanent】, the
default is 2 minutes. The steps to set the alarm pause time are as follows:
1. Select 【 Main Menu 】 quick keyfrom 【 System 】 column to select
【Maintenance】input maintenance passwordEnter.
2. Select【Alarm】submenu 【Pause /Reset】submenu.
3. Set【Alarm Pause Duration】.
7.7.2. Turn off all the Alarm
If 【Alarm Pause Duration】is set to 【Permanent】(Refer to section 7.7.1 Setting
Alarm Pause Time), you can press【Alarm Pause】quick key to turn off all alarms.
When
the alarm is turned off, it has the following characteristics:
No physiological alarm lamps flash and no physiological alarms are sounded.
The technical alarm sound is turned off, but the alarm light and alarm
information are still displayed
Display "Alarm Off" in the physiological alarm information area and the
background color is red.
Display alarm off icon in status area.
To exit the alarm off state, click【Alarm Pause】quick key
again.
WARNING:
Pausing or turning off the alarm may cause the patient to be in danger, please
handle it carefully.
7.8. Alarm Reset
Click on【Alarm Reset】quick key to reset the alarm system, and the alarm reset icon
will appear in the system status information area.