S10/S12/S10A/S12A Patient Monitor User’s Manual
19-9
Output
Parameters
Unit Full Name Formula
C(a-v)O2
mL/L,
mL/dL
arteriovenous
oxygen content
difference
C(a-v)O2=CaO2×CvO2
O2ER %
oxygen extraction
ratio
O2ER=CaO2CvO2/ CaO2
DO2 mL/min oxygen transport DO2=C.O.×CaO2
PAO2
mmHg,
kPa
partial pressure of
oxygen in the
alveolar
PAO2=FiO2×(ATMPwater
pressure) 】-(PaCO2×1.25)
Wherein the water pressure is
selected to be 47mmHg (6.3kPa)
AaDO2
mmHg,
kPa
alveolar-arterial
oxygen difference
AaDO2=PAO2PaO2
CcO2
mL/L,
mL/dL
capillary oxygen
content
CcO2 = Hb ×1.34
0.031×PAO2
Qs/Qt % venousad mixture
Qs/Qt= (CcO2CaO2)/( CcO2
CvO2)
19.6. Ventilation Calculations
The monitor provides the ventilation calculation function. The monitor can save the
results of up to 20 calculations, which are displayed in groups.
19.6.1. Calculation Step
The ventilation calculation steps are as follows:
1. Access ventilation calculation page by either of the following ways:
SelectCalculationsquick keyVentilationsubmenu
.
SelectMain Menuquick keyfrom Calculations column to

Table of Contents