S10/S12/S10A/S12A Patient Monitor User’s Manual
39
C.1.3 ST
Item Default Setting
ST-I, ST-II,
ST-III, ST-aVR,
ST-aVL,
ST-aVF, ST-V1,
ST-V2, ST-V3,
ST-V4, ST-V5,
ST-V6, ST-Va,
ST-Vb
Alarm switch ON
Alarm high limit 0.2 mV
Alarm low limit -0.2 mV
Alarm priority MED
Alarm print OFF
ST analysis OFF
ST marking line OFF
ST point automatic adjustment OFF
ST Point J + 60 ms
ISO Point -80 ms
J Point 48 ms
C.1.4 QT
Item Default Setting
QTc Alarm switch OFF
Alarm high limit 400ms
Alarm priority MED
Alarm print OFF
QTc Alarm switch OFF
Alarm high limit 40ms
Alarm priority MED
Alarm print OFF
QT analysis OFF