S10/S12/S10A/S12A Patient Monitor User’s Manual
21-3
Antenna: Build-in or External.
4. Restart the monitor
21.2.5. Connecting the Central Monitoring System (CMS)
The monitor can be connected to the central monitoring system via wired network or
wireless network.
21.2.5.1. Setting the CMS IP Address
To set the IP address of CMS, follow this procedure:
1. Select Main Menu quick keyfrom System column to select
Maintenanceinput maintenance passwordEnter.
2. Select Network Setup submenu CMS submenu.
3. Set the IP address of the CMS. The monitor can be received by the CMS
of the IP address.
21.2.5.2. Setting the device number of the monitor
The device number of the networked monitor will be displayed when the central
monitoring system and other beds are monitored. The steps for setting the device
number of the monitor are as follows:
1. Select Main Menu quick keyfrom System column to select
Maintenanceinput maintenance passwordEnter.
2. Select Network Setup submenu CMS submenu.
3. Set Device No. of the monitor.
Please refer to the Central Monitoring System User’s Manual for detailed
instructions.
NOTE: This monitor can only be connected to the central monitoring system
provided by the manufacturer. Do not try to connect the monitor to
other central monitoring system.