S10/S12/S10A/S12A Patient Monitor User’s Manual
44
Item Default Setting
Adult: 35 mmHg
Pediatric/Neonatal: 60mmHg
Low limit ART/Ao/UAP/BAP/FAP/LV/P1/P2 arterial pressure
Adult: 90 mmHg
Pediatric: 70 mmHg
Neonatal: 55 mmHg
PA/PAWP:
Adult: 10 mmHg
Pediatric/Neonatal: 24 mmHg
Alarm priority MED
Alarm print OFF
IBP-D Alarm switch ON
High limit ART/Ao/UAP/BAP/FAP/LV/P1/P2 arterial pressure
Adult: 90mmHg
Pediatric: 70 mmHg
Neonatal: 60mmHg
PA/PAWP
Adult: 16 mmHg
Pediatric/Neonatal: 4 mmHg
Low limit ART/Ao/UAP/BAP/FAP/LV/P1/P2 arterial pressure
Adult: 50 mmHg
Pediatric: 40 mmHg
Neonatal: 20 mmHg
PA/PAWP
Adult: 0 mmHg
Pediatric/Neonatal: -4 mmHg
Alarm priority MED
Alarm print OFF
IBP-M Alarm switch ON
High limit ART/Ao/UAP/BAP/FAP/LV/P1/P2 arterial pressure
Adult: 110 mmHg