S10/S12/S10A/S12A Patient Monitor User’s Manual
12-9
D, and E. You can set the measurement duration for each period and the interval between
NIBP measurements in each period separately. The steps to set up the NIBP measurement
sequence are as follows:
1. Select the NIBP parameter area a to enter theNIBPmenu.
2. SelectSequencesubmenu.
3. Set each sequence measurement toDurationorInterval Timeseparately.
12.8.7. Setting NIBP Unit
You can change the NIBP units by following these steps:
1. Select the NIBP parameter area a to enter theNIBPmenu.
2. SelectSetupsubmenu.
3. Set NIBPUnit.
12.8.8. Setting NIBP Invalid Time
NIBP measurements become outline fonts after a preset time. This avoids older NIBP
values being misinterpreted as current measurements. To set the timeout period, follow
this procedure:
1. Select Main Menu quick keyfrom System column to select
Maintenanceinput maintenance passwordEnter.
2. SelectModulesubmenuOthersubmenu.
3. SetNIBP Invalid Time.
12.8.9. Displaying the NIBP List
To display multiple sets of the latest NIBP measurements, follow this procedure:
1. EnterScreen Layoutsubmenu
by either of the following ways:
SelectScreen Setupquick keySelectScreen Layoutsubmenu.
SelectMain Menuquick keyfrom Display column to select
Screen Layout.
2. In the desired parameter area, selectNIBPNIBP
List.