S10/S12/S10A/S12A Patient Monitor User’s Manual
12-9
D, and E. You can set the measurement duration for each period and the interval between
NIBP measurements in each period separately. The steps to set up the NIBP measurement
sequence are as follows:
1. Select the NIBP parameter area a to enter the【NIBP】menu.
2. Select【Sequence】submenu.
3. Set each sequence measurement to【Duration】or【Interval Time】separately.
12.8.7. Setting NIBP Unit
You can change the NIBP units by following these steps:
1. Select the NIBP parameter area a to enter the【NIBP】menu.
2. Select【Setup】submenu.
3. Set NIBP【Unit】.
12.8.8. Setting NIBP Invalid Time
NIBP measurements become outline fonts after a preset time. This avoids older NIBP
values being misinterpreted as current measurements. To set the timeout period, follow
this procedure:
1. Select 【 Main Menu 】 quick keyfrom 【 System 】 column to select
【Maintenance】input maintenance passwordEnter.
2. Select【Module】submenu【Other】submenu.
3. Set【NIBP Invalid Time】.
12.8.9. Displaying the NIBP List
To display multiple sets of the latest NIBP measurements, follow this procedure:
1. Enter【Screen Layout】submenu
by either of the following ways:
Select【Screen Setup】quick keySelect【Screen Layout】submenu.
Select【Main Menu】quick keyfrom 【Display】 column to select
【Screen Layout】.
2. In the desired parameter area, select【NIBP】【NIBP
List】.