EasyManua.ls Logo

Biolight Meditech S10 - Page 133

Default Icon
307 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
S10/S12/S10A/S12A Patient Monitor User’s Manual
12-8
measurements. To set the NIBP interval, follow this procedure:
1. Select the NIBP parameter area or waveform area to enter theNIBPmenu.
2. SelectSetupsubmenu.
3. SetInterval Time.
12.8.4. Selecting NIBP Start Mode
Start mode defines how NIBP auto mode works. To set the Start Mode, follow
this procedure:
1. Select the NIBP parameter area a to enter theNIBPmenu
2. SelectSetupsubmenu.
3. SetStart Mode.
Clock: After the first measurement, the monitor automatically
synchronizes NIBP automatic measurements with the real time clock.
For example, if Interval is set to 30min, and you start NIBP auto
measurement at 10:03, the next measurement will be taken at 10:30, and
then at 11:00, 11:30, and so on.
Interval: After the first measurement, the monitor automatically
repeats measurements at set interval. For example, if Interval is set
to30min, and you start NIBP auto measurement at 10:03, the next
measurement will be taken at 10:33, and then at 11:03, 11:33, and so on.
12.8.5. Enabling the NIBP End Tone
The monitor can issue a reminder tone at the completion of NIBP measurement. The
NIBP End Tone is off by default. To switch on the NIBP end tone, follow this procedure:
1. Select the NIBP parameter area a to enter theNIBPmenu
2. SelectSetupsubmenu.
3. Switch onNIBP
End Tone.
12.8.6. Setting NIBP Sequence Measurement
NIBP sequence measurements can consist of up to 5 measurement periods: A, B, C,

Table of Contents