S10/S12/S10A/S12A Patient Monitor User’s Manual
15-9
2. Select【Alarm】submenu.
If the alarm setting is protected by password, enter the password. For
detail, please refer to 7.6.2 Changing Alarm Setup Protection Mode.
3. Set alarms as needed
15.6.2. Settings Apnea Alarm Time
You can set the apnea alarm time by following the steps below. The monitor will
trigger the alarm when the patient’s suffocation time exceeds the set time.
1. Select the CO
2
parameter area or waveform area to enter the【CO
2
】menu.
2. Select【Alarm】submenu.
3. Set【Apnea Delay】.
WARNING:
It is cannot judge the cause of respiratory apnea through CO
2
monitoring. If the
monitor cannot detect the patient's breath again from the moment of the last
breath to the preset time, the monitor only provides the alarm of respiratory
apnea. Therefore, the alarm of respiratory apnea should not be used for the
diagnosis of the patient.
15.6.3. Changing the CO
2
Unit
To change the CO
2
unit, follow this procedure:
1. Select the CO2 parameter area or waveform area to enter the【CO
2
】menu.
2. Select【Setup】submenu.
3. Set CO
2
【Unit】.
15.6.4. Setting the CO
2
Waveform
To set the CO
2
waveform, follow this procedure:
1. Select the CO2 parameter area or waveform area to enter the【CO
2
】menu.
2. Select【Setup】submenu.
3. Set CO
2
waveform【Wave Mode】, 【Wave Speed】or【Scale】.