S10/S12/S10A/S12A Patient Monitor User’s Manual
28
Alarm limit Range Step
Neonatal: (low limit+1 mmHg) ~110 mmHg
NIBP-D-Low 10 mmHg ~ (high limit-1 mmHg)
A.9.4 SpO2
Alarm limit Range Step
SpO2 High
(low limit+1) ~100
1
SpO2 Low
(SpO2 Desat +1) ~ (high limit-1)
SpO2 Desat
0~ (low limit-1)
PR High (PR low limit+1bpm)~350bpm
1bpm
PR Low 0bpm~ (PR high limit-1bpm)
A9.5 TEMP
Alarm limit Range Step
T1/T2 High (low limit+0.1ºC) ~50.0ºC 0.1 ºC
T1/T2 Low 0 ºC~ (high limit-0.1ºC) 0.1 ºC
TD High 0ºC~5.0ºC 0.1 ºC
A.9.6 IBP
Alarm limit Range Step
IBP-M-High (low limit +1mmHg) ~360mmHg
1mmHg
IBP-M-Low -50mmHg~ (high limit -1mmHg)
IBP-D-High (low limit +1mmHg) ~360mmHg
1mmHg
IBP-D-Low -50mmHg~ (high limit -1mmHg)
NIBP-S-High (low limit +1mmHg) ~360mmHg
1mmHg
NIBP-S-Low -50mmHg~ (high limit -1mmHg)