S10/S12/S10A/S12A Patient Monitor User’s Manual
56
SpO2 Sensor
Disconnected
Low
Sound and light
can be cleared
The SpO2 main cable falls off
from the module end or the
connection between the SpO2
sensor and the SpO2 main cable
falls off. Confirm that SpO2 main
cable and sensor are connected
normally. If the alarm still cannot
be eliminated, replace the sensor.
SpO2 Low
Confidence
Low Not clearable PI<0.3% or signal quality <60.
SpO2 Update
Timeout
Low Not clearable 25s SpO2 measurement data not
updated.
SpO2 Motion
Interference
Low Not clearable Patients move too much, affecting
measurement.
SpO2 Searching
Pulse
Prompt
/ SpO2 module is searching for
pulse.
SpO2 Sensor
Incompatible
Med Not clearable
Non-factory SpO2 sensor are
used. Please replace the original
sensor.
SpO2 Sensor Has
Expired
Med Not clearable
SpO2 sensor has expired. Please
replace the sensor.
SpO2 Sensor is
About to Expire
Prompt /
SpO2 sensor is about to expire.
Please replace the sensor in time.
D.2.5 TEMP Technical Alarm
Alarm messages
Default
priority
Alarm clear
method
Cause and solution
TEMP Self-test
Error
Med Not clearable
Board failure. Please contact the
manufacturer for repair.
TEMP label
Sensor Off
Med
Completely
clear
Check the connection of the sensor and
reconnect the sensor.