S10/S12/S10A/S12A Patient Monitor User’s Manual
59
Alarm messages
Default
priority
Alarm clear
method
Cause and solution
sensor is not connected or the
pressure is out of range or the
pressure is unstable.
XX Catheter Off HIGH Not clearable
The catheter is pulled out from the
patient. Please check the
connection.
Zero Required Prompt / /
Zero Succeed Prompt /
The IBP module is zeroing
successful
Note: XX represents IBP labels, such as PA, CVP, FAP, P1, etc.
D.2.8 C.O. Technical Alarm
Alarm messages
Default
priority
Alarm clear
method
Cause and solution
C.O. TB Self-test
Error
Med Not clearable
The board is failure. Please
contact the factory for repair.
C.O. TI Self-test
Error
Med Not clearable
The board is failure. Please
contact the factory for repair.
C.O. TB Sensor Of
f
Med Completely clear
Check the sensor connection and
reconnect the sensor.
C.O. T1 Sensor Off Med Completely clear
Check the sensor connection and
reconnect the sensor.
C.O. Measure
Timeout
Low Completely clear
When measuring manually, it do
not inject for sustaining 20
seconds.