S10/S12/S10A/S12A Patient Monitor User’s Manual
7-5
7.6.2. Changing Alarm Setup Protection Mode
You can change the password protection mode of the alarm settings and arrhythmia
settings as follows:
1. Select Main Menu quick keyfrom System column to select
Maintenanceinput maintenance passwordEnter.
2. SelectAuthorizationsubmenu.
3. Change the password protection mode of the alarm settings.
No Password: Change alarm setups to not be password protected.
Password: Change alarm switch, alarm limit and alarm level to be
protected by password.
If you use password to access alarm and arrhythmia alarm related settings, you can
set the valid time of the password, beyond which you need to re-enter the password. For
details, please refer to 5.9 Setting Password Valid Time.
7.6.3. Setting Alarm Sound Properties
7.6.3.1. Setting Alarm Volume
1. EnterSetupinterface in either of the following ways:
SelectAlarm Setupquick keySelectSetupsubmenu.
SelectMain Menuquick keyfromAlarmcolumn to select Setup.
2. Set Alarm Volume. The alarm volume range is X-10, in which X is the
minimum volume, depending on the set minimum alarm volume, and 10 is the
maximum volume.
3. SetHigh Alarm Volume.
4. SetReminder Volume.
NOTE
When the alarm volume is set to 0, the alarm tone will be turned off, and an
alarm audio off icon will appear on the screen.
When the alarm volume is set to 0, the setting of high level alarm volume is
invalid.