S10/S12/S10A/S12A Patient Monitor User’s Manual
7-7
High Alarm Interval
:
3
15s, and the default value is 10s.
Medium Alarm Interval
: 3
30s,
and the default value is 20s.
Low Alarm Interval
: 16
30s
and the default value is 20s.
7.6.3.5. Setting Reminder Volume
When the alarm volume is 0, alarm reset or the alarm is off, the monitor can provide
periodic alarm prompt tone to remind you that there is still an activated alarm in the
current system. This function is turned on by default.
You can set the alarm tone as follows:
1. Select Main Menu quick keyfrom System column to select
Maintenanceinput maintenance passwordEnter.
2. SelectAlarmbuttonPause /Resetsubmenu.
3. Set Alarm Pause duration. You can set Alarm Pause durationto
1min, 2min, 3minorPermanent, the default is2min.
4. SetAlarm Off Reminderswitch.
On: The monitor provides an alarm tone according to the set interval.
Off: The monitor does not provide an alarm tone.
5. Set Reminder Interval. You can set
Reminder Intervalto 1min,
2min, 3min, 5minor10min, the default is5min.
7.6.3.6. Setting Alarm Tone Enhancement
The monitor provides an alarm tone enhancement function. If the alarm exceeds the
set time and is not confirmed, the alarm volume can be automatically enhanced.
The steps to set the alarm tone enhancement are as follows
1. Select Main Menu quick keyfrom System column to select
Maintenanceinput maintenance passwordEnter.
2. SelectAlarmSound.
3. SetAuto Increase Volumeto3 Steps, 2 Steps, 1 Stepsor
Off.
3 Steps:
After the alarm occurs, the alarm volume will be