S10/S12/S10A/S12A Patient Monitor User’s Manual
7-7
【
High Alarm Interval
】
:
3
~
15s, and the default value is 10s.
【
Medium Alarm Interval
】
: 3
~
30s,
and the default value is 20s.
【
Low Alarm Interval
】
: 16
~
30s
,
and the default value is 20s.
7.6.3.5. Setting Reminder Volume
When the alarm volume is 0, alarm reset or the alarm is off, the monitor can provide
periodic alarm prompt tone to remind you that there is still an activated alarm in the
current system. This function is turned on by default.
You can set the alarm tone as follows:
1. Select【 Main Menu 】 quick keyfrom 【 System 】 column to select
【Maintenance】input maintenance passwordEnter.
2. Select【Alarm】button【Pause /Reset】submenu.
3. Set 【Alarm Pause duration】. You can set 【Alarm Pause duration】to
【1min】, 【2min】, 【3min】or【Permanent】, the default is【2min】.
4. Set【Alarm Off Reminder】switch.
【On】: The monitor provides an alarm tone according to the set interval.
【Off】: The monitor does not provide an alarm tone.
5. Set 【Reminder Interval】. You can set
【Reminder Interval】to 【1min】,
【2min】, 【3min】, 【5min】or【10min】, the default is【5min】.
7.6.3.6. Setting Alarm Tone Enhancement
The monitor provides an alarm tone enhancement function. If the alarm exceeds the
set time and is not confirmed, the alarm volume can be automatically enhanced.
The steps to set the alarm tone enhancement are as follows
1. Select 【 Main Menu】 quick keyfrom【 System 】 column to select
【Maintenance】input maintenance passwordEnter.
2. Select【Alarm】【Sound】.
3. Set【Auto Increase Volume】to【3 Steps】, 【2 Steps】, 【1 Steps】or
【Off】.
【3 Steps】:
After the alarm occurs, the alarm volume will be