EasyManua.ls Logo

Biotage Isolera Four - Replace the Detector Lamp(S)

Default Icon
116 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
411829-L, Maintenance April 2012 Page 5-11
5
4. Mount the needle on the collection arm. Ensure that the needle is touching the needle guide.
Figure 5-11. Mount the Needle on the Collection Arm
5.13 Replace the Detector Lamp(s)
1. Shut down the system as described on page 4-1.
2. Disconnect the power cord from the power outlet.
3. Remove the internal detector service panel at the left side of the system using a Torx 10
screwdriver.
4. Loosen the screw locking the UV lamp in a vertical position (see Figure 5-12) using a Torx 20
screwdriver.
Figure 5-12. Loosen the Locking Screw
(Fixed and Variable Detector Shown)
5. Turn the UV lamp into a horizontal position with the lamp socket facing you; see Figure 5-13.
6. Remove the two screws on the lamp socket (see Figure 5-13) using a Torx 10 screwdriver.
Figure 5-13. Turn the UV Lamp and Remove
Mount the
needle here
The needle
should touch the
needle guide
Locking screw
Lamp socket

Table of Contents

Related product manuals