The following softkeys will be displayed:
ON: Press this soft key to enable the open/short/load Correction data.
OFF: Press the soft key to disable the open/short/load Correction data.
MEAS OPEN: Press this soft key to execute open correction at the frequency
entered in step number 8.
MEAS SHORT: Press this soft key to execute short correction at the specified
frequency.
MEAS LOAD: Press this soft key to execute the load correction at FREQ selected in
step number 8.
9. Connect the test fixture to the test terminal.
OPEN CORRECTION
a. Disconnect any device under test from the test leads or test fixture and make
sure the test terminals are open (nothing connected).
b. Press MEAS OPEN to perform open correction at the current set frequency. The
test result (G, B) of the open correction test will be displayed at the bottom of
the screen.
c. Move the cursor to OPEN.
d. Press ON to perform an open correction calculation at the selected frequency in
later measurements.
SHORT CORRECTION
a. Disconnect any device under test from the test leads or test fixture and make
sure the test terminals are open (nothing connected).
b. Press MEAS SHORT to perform short correction at the current set frequency. The
test result (G, B) of the open correction test will be displayed at the bottom of
the screen.
c. Move the cursor to SHORT.
d. Press ON to perform a short correction calculation at the selected frequency in
later measurements.
LOAD CORRECTION
a. Using the cursors, navigate to REF A field.
b. Enter a known value or reference value (for the component to be used as a
standard) for the primary parameter set in the FUNC field.
c. Using the cursors, navigate to REF B field.
Enter a known value or reference value for the secondary parameter set in the
FUNC field (for the component to be used as a standard).
10. Move the cursor to the FREQ display field.
11. Connect the component to be used as a standard to the test fixture.
12. Press MEAS LOAD, and the instrument will execute a load correction. The real test
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the BK Precision 894 and is the answer not in the manual?

BK Precision 894 Specifications

General IconGeneral
BrandBK Precision
Model894
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals