TT100 In-Line Viscometer Installation, Operation, Maintenance and Service Instructions
P/N M/97-520 3-5
Figure 3-3:Typical Calibration after Component Replacement
START
Perform Component
Replacement Procedure(s)
(Refer to Section 6)
Perform Single or Dual
Range Viscometer
Span Calibration
(Refer to Page 3-10 or 3-14)
END
Operate
Viscometer
(Refer to Section 4)
Perform Coarse Zero
Calibration
(Refer to Page 3-18)
Perform Fine Zero Voltage
Calibration
(Refer to Page 3-21)
Were armature and
Microsyn field coil
disturbed during
component replacement?
YES
NO
Perform Preliminary
Torque Sensor Test
(Refer to Page 3-8)