TT100 In-Line Viscometer Installation, Operation, Maintenance and Service Instructions
P/N M/97-520 5-7
Figure 5-3: Stator Installation and Stop Post Centering
NOTE: If the stop posts will not align in the center of the stator holes,
remove the mounting screws and rotate the stator 180º. Install the screws
and repeat step 6. The stop posts may also be bent slightly to insure correct
alignment. Refer to Appendix A and contact Brookfield Engineering
Laboratories if the stop posts cannot be properly aligned.
7. Place the sensing endcap assembly in a safe place and proceed with
Cleaning the Drive Endcap Assembly and Measuring Chamber.
Cleaning the Drive Endcap Assembly and Measuring Chamber
In order to remove and clean the drive endcap, the drive motor, gearbox,
and mounting bracket must be removed from the measuring chamber.
They can be removed as one unit as shown in Figure 5-1.
CAUTION
Do not remove the drive endcap assembly as a unit unless the sensing end-
cap has been removed first.
1. Note the orientation of the drive motor bracket and measuring
chamber. The bracket should be reinstalled in the same position
during assembly.
CORRECT
INCORRECT
STOP
POSTS
MOUNTING
SCREWS