Operation
4-4 P/N M/97-520
Shutdown
CAUTION
Do not allow the process fluid to dry or solidify. Damage to the TT100
In-Line Viscometer may result.
1. Shut the TT100 In-Line Viscometer OFF by performing the
following steps:
a. Set the POWER switch on the Torque Sensor Electronics
Enclosure to the OFF position.
b. Set the POWER switch on the Motor Control Enclosure to
the OFF position.
c. If equipped with a double mechanical seal assembly, release
the pressure on the pressurized seal fluid.