TT100 In-Line Viscometer Installation, Operation, Maintenance and Service Instructions
P/N M/97-520 6-11
m. Remove the range jumper from the old circuit board and
connect it to the same terminals on J2.
9. Refer to Figure 3-3 within Section 3 - Calibration.
Torsion Element Replacement
WARNING
Make sure power has been turned OFF at the motor and torque sensor
electronics enclosures and the viscometer is empty and dry before ser-
vicing the viscometer.
1. Shut power OFF to the motor and torque sensor electronics
enclosures.
2. Remove seven of the eight screws that secure the sensing endcap to
the measuring chamber.
3. While holding the sensing endcap in position, remove the last screw
and carefully remove the sensing endcap from the measuring
chamber.
4. Remove the two mounting screws from the top of the stator as
shown in Figure 6-6.
Figure 6-6:Stator Removal
STOP POST
STOP POST
STATOR
STATOR MOUNTING
SCREWS
TORSION ELEMENT WIRES
NOTE: There are many different versions of the sensing endcap. Your viscometer may
look different from the viscometer shown.