Service
6-10 P/N M/97-520
Figure 6-5:Torque Sensor Electronics Enclosure Signal and Power Connection Locations
8. Make the following electrical connections:
a. Connect the L1 power wire to J3-T1.
b. Connect the L2 power wire to J3-T2.
c. Connect a GROUND JUMPER wire between the cover
ground stud and J3-T3.
d. Connect the RED sensor wire to J1-T11.
e. Connect the GREEN sensor wire to J1-T10.
f. Connect the PURPLE sensor wire to J1-T9.
g. Connect the BLUE sensor wire to J1-T8.
h. Connect the YELLOW sensor wire to J1-T7.
i. Connect the GROUND power wire to the ground stud.
j. Connect the SHIELD wire to J1-T12.
k. Connect the wire connectors to the Microsyn field coil as
shown in Figure 6-5.
l. Connect the display device to J1-T1 and T2.
1243
12
11
10
9
8
7
6
5
4
3
2
1
1234 123
PRIMARY - (RED)
PRIMARY + (GREEN)
SECONDARY + (YELLOW)
SECONDARY - (BLUE)
L1
L2
GROUND
SHIELD
+
J3
J2
J1
MOUNTING SCREW HOLES
GROUND - (PURPLE)