TT100 In-Line Viscometer Installation, Operation, Maintenance and Service Instructions
P/N M/97-520 3-17
24. Turn power OFF to the torque sensor electronics enclosure.
25. Remove the DVM wire connections from the torque sensor
electronics circuit board.
26. Reconnect the milliampere cable to J1 Pins 1 and 2 on the torque
sensor electronics circuit board and install the enclosure cover.
27. Remove the sensing endcap from the vise and careful install it into
the viscometer.
28. Install the sensing endcap mounting bolts and tighten them in a
rotational pattern.
29. Turn power ON to the torque sensor electronics enclosure.
CAUTION
Do not run the motor without fluid in the viscometer. Damage to the
mechanical shaft seals may result.