34 Chapter 5 Acquiring Panoramic Images
Launching the X-ray
To launch the X-ray, follow these steps:
1. Leave the X-ray room and close the door. You must keep visual contact with the patient during
acquisition.
2. Launch the X-ray with the remote control.
The turns yellow and a warning sound is heard, indicating X-ray emission.
Press and hold the exposure button until the end of the acquisition is indicated by the on-screen
“Release Handswitch” message.
When the acquisition ends, the Acquisition interface disappears and the acquired image is
automatically transferred to the Imaging Window.
3. Check the image quality.
4. Do the following when the acquisition is finished:
• Release the patient.
• Remove the hygiene barrier from the bite block.
Important: To stop the acquisition, if there is any problem, release
the exposure button of the remote control or press the red
emergency stop button.