Maintenance 8-7
8.4.3 Running Scanner Maintenance - Steps
The steps for running Scanner Maintenance described below are the same regardless of which of the
configuration solutions you use.
To run Scanner maintenance Program:
On Scanners: FLEX50i
1. Power on the scanner if not already ON.
2. Insert the Calibration Sheet that came with your scanner. The sheets printed side must be
inserted face down.
3. Start the Scanner Maintenance program in one of the ways described at the start of this section.
4. The program will run calibration.
5. After calibration, remove the Calibration Sheet from the scanner.
6. Return the Scanner Calibration Sheet to its protective cover and then place it in the storage folder.
Store the folder in a dry place and out of direct light.
On Scanners: HD4230i
1. Power on the scanner if not already ON.
2. Insert the Calibration Sheet that came with your scanner. Align the midpoint arrow with the middle
mark on the scanner. The sheets printed side must be inserted face down.
3. Start the Scanner Maintenance program in one of the ways described at the start of this section.
4. The program will run Camera Alignment, Stitching and Calibration.
5. The program analyses alignment and stitching status of the cameras. If the cameras are unaligned,
the program and scanner will automatically make the necessary corrections.
6. After Camera Alignment, the Scanner Maintenance program automatically moves on to Calibration.
7. Calibration completes the maintenance process. Remove the Scanner Calibration Sheet from the
scanner.
8. Return the Scanner Calibration Sheet to its protective cover and then place it in the storage folder.
Store the folder in a dry place and out of direct light.