1 Connect the nebuliser hose to the port on the front
panel of the Evita 4.
Fill the medicament nebuliser in accordance with the
specific Instructions for Use.
Warning: the effect of aerosols on sensors, filters
and heat and moisture exchangers (HME) must be
taken into account.
The measuring function of the flow sensor may be
impaired.
The flow resistance of filters is liable to increase and
may impair ventilation.
Do not place a microbial filter on the nebuliser outlet
during nebulisation.
During medicament nebulisation, do not use a heat
and moisture exchanger (HME) at the Y-piece.
Risk of increased breathing resistance!
2 Hold down » « key until the yellow LED lights up.
Note message on-screen:
Nebuliser on!
The nebuliser operates for 30 minutes.
If medicament nebulisation needs to be stopped
prematurely:
2 Press » « key again. The yellow LED goes out,
and the nebuliser is switched off.
The flow sensor is then automatically cleaned and
calibrated.
Screen display:
Flow Calibration
Remove any left-over medicament. Follow the Instruc-
tions for Use of the medicament nebuliser.
90
Operation
Special Functions
Medicament nebulisation
D Evita 4
1
D Evita 4
2

Table of Contents

Other manuals for Dräger Evita 4

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Dräger Evita 4 and is the answer not in the manual?

Dräger Evita 4 Specifications

General IconGeneral
BrandDräger
ModelEvita 4
CategoryMedical Equipment
LanguageEnglish

Related product manuals