100-412-183 REV. 13 109
3
Go to the P/S Diagnostics tab. Press the Start Test button to start the test.
Sonics will become active and the button will change to Stop Test. Press the
button again to stop the test.
If the OK - Memory Stored indicator becomes active the test procedure is
finished.
NOTICE
Low level ultrasonic energy will remain active until the Stop Test button is
pressed. If you close the web browser, ultrasonic energy will turn off
automatically after 5 seconds.
4
If the Overload - Memory Cleared indicator becomes active the test will be
interrupted, press the Reset Overload button and repeat step 3 one time
only. If the alarm persists, refer to 8.6 Troubleshooting.
Table 7.12 Power Supply Ultrasonic Test Procedure (Web Page Interface)