5.3 Installation Steps
Basic installation notes:
To avoid problems associated with EMI, you should route high power lines (AC and Ultrasonic RF)
away from low power lines (controls signals)
You should consider future troubleshooting and repair when installing all wiring. All wiring should
be either color coded or tagged with industrial wire tags
The minimum cable bend radius is 5 times the cable outer diameter for RF cables
The minimum cable bend radius is 10 times the cable outer diameter for user I/O & Ethernet
cables
Ground wires should not be shared with other equipment
All inductive coils must be suppressed with appropriate devices, such as diodes or RC networks
WARNING High Voltage Hazard
To prevent the possibility of an electrical shock:
Ensure the power source is disconnected before beginning work on line
connections
Ensure the power switch on the back of the unit is in the OFF position
before making any electrical connections
Always plug the power supply into a grounded power source
To prevent the possibility of an electrical shock, ground the power supply
by securing an 8 gage grounded conductor to the ground screw located
next to the air outlet
Ensure power supply installation is performed by qualified personnel and
in accordance with local standards and regulations