DL8000 Preset Instruction Manual
Revised February-2016 Communications Protocols D-29
Clear Display
[0x38]
Flags Immediately Set
None
Flags Immediately
Cleared
None
Exceptions
Passcode Entry in Progress
Operating Mode is Manual
K&D Locked Out
Supporting
Parameters
None
Results
Clear existing screen and display “Please wait”
until next command is received. The alarm and
date-time continue to display.
Report Alarm
[0x3D]
Raises alarms based on reset alarm ID [63,0,173]. Possible uses of this
command are:
Raising an alarm based on conditions that are either not input to or not
detectable by the DL8000 (such as a fire detection system or a PLC being
used to monitor a rail car’s position).
Stopping a batch with a situation-specific message. In this case, the TAS
might configure a safety circuit alarm message via communications and
then raise the corresponding safety circuit alarm. The TAS can reset the
alarm using the reset primary alarms command [0x14].
Note: A TAS should not depend on DL8000 alarms to prevent or control
hazardous or life-threatening conditions.
Flags Immediately Set
Primary Alarm (if primary alarm is raised)
Flags Immediately
Cleared
None
Exceptions
Invalid alarm ID
Unable to set or reset alarm
Supporting
Parameters
Reset alarm ID [63,0,173]
Results
Batch may stop due to reported alarm.